<3. INSTALLATION>

3-8

IM 01F06A00-01EN

Table 3.3 (a) Installation of Wafer Type Vortex Flowmeter

Wafer type

Description

When Installation Collars are not required,the
installation vortex flowmeters applied to the
following line sizes and flanges.

Size mm (inch)

Flange Rating

50(2)

80(3)

100(4)

JIS 20K, 40K
ANSI class 300,600

JIS 10K, 20K, 40K
ANSI class 300, 600

JIS 10K, 20, 40K
ANSI class 150, 300, 600

Vertical Installation

Horizontal Installation

(1) Insert two stud bolts in the bolt holes

on the flowmeter shoulder to align
the instrument body with the inner
diameter of the adjacent piping.


(2) Tighten all bolts uniformly and check

that there is no leakage between the
instrument and the flanges.

F0312.ai

Flow Direction

Nut

Nut

Gasket

Bolt Hole

Gasket

Flange

Flange

Flow
Direction

Stud Bolt (8 pcs.)

Electrical Connection

Electrical

Connection

Electrical
Connection

Flange

Flow
Direction

Nut

Gasket

Collar (4 pcs.)

Collar (4 pcs.)

Gasket

Flange

Gasket

Gasket

Flange

Flange

Bolt (4 pcs.)

Bolt (4 pcs.)

Nut

Nut

Nut

Electrical
Connection

Flow Direction

Size mm (inch)

Flange Rating

15 to 40

(1/2 to 1-1/2)

50(2)

80(3)

All ratings

JIS 10K, ANSI class 150,
DIN PN10 to PN40

ANSI class 150

Vertical Installation

When Installation Collar are required, the
installation vortex flowmeters applied to the
following line sizes and flange ratings.

The inside diameter of the gasket must
be larger than the pipe inner diameter
so that it will not disturb the flow in the
pipeline.

Horizontal Installation

When installing the Flowmeter vertically in the
open air, change the electrical connection port
direction to the ground. If the electrical
connection port is installed upwards, rain
water might leak in.

WARNING

In case of vertical installation, two collars in
the upper part might move after the
installation. But it doesn't influence the
performance, please use the flowmeter under
such condition.

WARNING

WARNING

(1) Insert two collars on each two bolts of bottom side of the flowmeter.
(2) Fit the flowmeter body to the collars. And tighten the four bolts and nuts

uniformly.

(3) Check for leakage from the flange connections.