Installation, Reducer pipe, Expander pipe – Yokogawa digitalYEWFLO (DY) User Manual

Page 12: Bent pipe and straight pipe length, Valve position and straight pipe length

Advertising
background image

<3. INSTALLATION>

3-2

IM 01F06A00-01EN

Table 3.1 (a) Straight pipe length and recommendations (1)
D: Nominal diameter (mm)

Description

Figure

Reducer pipe:

Ensure the upstream straight pipe length to be 5D or more, and the
downstream straight pipe length to be 5D or more for per reducer
pipe.

digitalYEWFLO

Flow

5D or more

Reducer

5D or more

Expander pipe:

Ensure the upstream straight pipe length to be 10D or more, and
the downstream straight pipe length to be 5D or more for per
expander pipe.

digitalYEWFLO

Flow

10D or more

Expander

5D or more

Bent pipe and straight pipe length:

1. Single bent pipe

2. Double bent pipe; coplanar

3. Double bent pipe; non coplanar

1.

2.

3.

digitalYEWFLO

Flow

10D or more

5D or more

digitalYEWFLO

Flow

10D or more

5D or more

digitalYEWFLO

Flow

20D or more

5D or more

Valve position and straight pipe length:

 Install the valve on the downstream side of the flowmeter.

The upstream straight pipe length dependent on the element
located on the upstream such as reducer/expander, bent and etc.,
refer to description as above. Keep 5D or more for downstream
straight pipe length.

 In case the valve has to be installed on the upstream of the

fl owmeter, ensure the upstream straight pipe length to be 20D or
more, and the downstream straight pipe length be 5D or more.

Refer to each element above for

straight pipe run.

digitalYEWFLO

Flow

digitalYEWFLO

Valve

5D or more

20D or more