7) item h : adjust, 8) item j : test – Yokogawa digitalYEWFLO (DY) User Manual

Page 45

Advertising
background image

<6. PARAMETERS>

6-9

IM 01F06A00-01EN

(7) Item H : Adjust.

These items are for setting of adjustment.

A value in “( )” is the data corresponding to the indicator.

Item

Name

R / W

Data Range

Unit

Decimal

Point

Remark

Initial
Value

Disp. U / D

(*1)

H00 ADJUST

Menu H (Adjust)

H10 TRIM 4mA

W -1.00 to 1.00

%

2

Trimming 4mA

0.0

D

H11 TRIM 20mA

W -1.00 to 1.00

%

2

Trimming 20mA

0.0

D

H20 USER ADJUST

W 0.00001 to 32000

0 to 5

User Adjust

1.0

D

H25 REYNOLDS ADJ

W NOT ACTIVE

(0)

ACTIVE (1)

Reynolds Coeffi cient

(0)

D

(Indicate and Set only for H25: ACTIVE)

H26 DENSITY f

W 0.00001 to 32000

D25

0 to 5

Density at operating condition

1024

D

H27 VISCOSITY

W 0.00001 to 32000

mPa.s

0 to 5

Viscosity factor

1.0

D

H30 EXPANSION FA

W NOT ACTIVE

(0)

ACTIVE (1)

Expansion correction for
compressible Gas

(0)

D

H40 FLOW ADJUST

W NOT ACTIVE

(0)

ACTIVE (1)

Instrumental Error Adjust

(0)

D

(Indicator and Set only for H40: ACTIVE)

H41 FREQUENCY 1

W 0 to 32000

Hz

0 to 5

First break-point frequency

(f1)

0.0

D

H42 DATA 1

W -50.00 to 50.00

%

2

First correcting value

(d1)

0.0

D

H43 FREQUENCY 2

W 0 to 32000

Hz

0 to 5

Second break-point frequency (f2)

0.0

D

H44 DATA 2

W -50.00 to 50.00

%

2

Second correcting value

(d2)

0.0

D

H45 FREQUENCY 3

W 0 to 32000

Hz

0 to 5

Third break-point frequency

(f3)

0.0

D

H46 DATA 3

W -50.00 to 50.00

%

2

Third correcting value

(d3)

0.0

D

H47 FREQUENCY 4

W 0 to 32000

Hz

0 to 5

Fourth break-point frequency

(f4)

0.0

D

H48 DATA 4

W -50.00 to 50.00

%

2

Fourth correcting value

(d4)

0.0

D

H49 FREQUENCY 5

W 0 to 32000

Hz

0 to 5

Fifth break-point frequency

(f5)

0.0

D

H50 DATA 5

W -50.00 to 50.00

%

2

Fifth correcting value

(d5)

0.0

D

H60 SELF CHECK

R

GOOD
ERROR

Self-diagnostic message

(*1) : Available for 3.10 or greater version that can be checked in K50.

(8) Item J : Test

These items are for test of output.

A value in “( )” is the data corresponding to the indicator.

Item

Name

R / W

Data Range

Unit

Decimal

Point

Remark

Initial
Value

Disp. U / D

(*1)

J00 TEST

Menu J (Test)

J10 OUT ANALOG

W 0.0 to 110.0

%

1

Current Output

0.0

D

J20 OUT PULSE

W 0 to 10000

PPS

0

Pulse Output

0

D

J30 OUT STATUS

W OFF

(0)

ON (1)

Status Output

(0)

D

J40

(*2)

RELEASE TIME

W 10min

(0)

30min (1)
60min (2)
3h

(3)

6h

(4)

12h (5)

Test auto release time

0

D

J60 SELF CHECK

R

GOOD
ERROR

Self-diagnostic message

(*1) : Available for 3.10 or greater version that can be checked in K50.
(*2) : Available for 7.00 or greater version that can be checked in K50.