4.2.2.93. Setup.SendMeas.Assembly.CyclNo ON, OFF

Setup.SendMeas.Assembly.I ON, OFF

Setup.SendMeas.Assembly.Meas ON, OFF

Setup.SendMeas.Assembly.Pot ON, OFF

Setup.SendMeas.Assembly.IPulse ON, OFF

Setup.SendMeas.Assembly.Bur.V ON, OFF

Selection of the values from Assembly for the output in the set time interval

(see 4.2.2.92):

.CyclNo: Cycle number of the potential measurement. Together with the

cycle time (4.2.2.77), a time frame can be set up.

The cycle number is set to 0 on switching on the instrument and it

is always incremented as long as the instrument remains

switched on.

.I:

Total charge in mA·s associated to the cycle number, e.g.

"667.48".

.Meas:

Measured value in mV associated to the cycle number, e.g.

"104.2".

.Pot:

Voltage at generator electrode associated to the cycle number.

0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3

means >28 V.

.IPulse:

Current of pulse associated to the cycle number.

1 means 100 mA, 2 means 200 mA, 3 means 400 mA.

.Bur.V:

Dosed volume of connected Dosino in ml, e.g. "5.234".

The unit "assembly" must be preset (see 4.2.2.92).

4.2.2.94. Setup.SendMeas.Titrator.CyclNo ON, OFF

Setup.SendMeas.Titrator.Water ON, OFF

Setup.SendMeas.Titrator.Meas ON, OFF

Setup.SendMeas.Titrator.dWaterdt ON, OFF

Setup.SendMeas.Titrator.I ON, OFF

Setup.SendMeas.Titrator.Pot ON, OFF

Setup.SendMeas.Titrator.IPulse ON, OFF

Selection of the values from the titrator which are sent in the set time interval

(see 4.2.2.91):

.CyclNo: Cycle number. Together with the cycle time (4.2.2.78), a time

frame can be set up. The other data belong to the corresponding

cycle number. The cycle number is set to 0 at the start of a

method and it is incremented until the end of the method.

.Water:

Total water associated to the cycle number in ug, e.g. "62.313"

.Meas:

Measured value in mV at the indicator electrode associated to the

cycle number, e.g. "104.2".

.dWaterdt: Rate or drift associated to the cycle number in ug/min, e.g. "23.0".

.I:

Total charge in mA·s associated to the cycle number, e.g.

"667.48".

.Pot:

Voltage at generator electrode associated to the cycle number.

0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3

means >28 V.

.IPulse:

Current of actual pulse associated to the cycle number.

1 means 100 mA, 2 means 200 mA, 3 means 400 mA.

OV will be sent for "overrange".

The unit "titrator" must be preset (see 4.2.2.92).