4.2 Remote control commands

756/831 KF Coulometer, Instructions of Use

93

4.2.2.95. Setup.AutoInfo.Status ON, OFF

Setup.AutoInfo.P ON, OFF

Setup.AutoInfo.T.R ON, OFF

Setup.AutoInfo.T.G ON, OFF

Setup.AutoInfo.T.GC ON, OFF

Setup.AutoInfo.T.S ON, OFF

Setup.AutoInfo.T.B ON, OFF

Setup.AutoInfo.T.F ON, OFF

Setup.AutoInfo.T.E ON, OFF

Setup.AutoInfo.T.O ON, OFF

Setup.AutoInfo.T.N ON, OFF

Setup.AutoInfo.T.Re ON, OFF

Setup.AutoInfo.T.Si ON, OFF

Setup.AutoInfo.T.M ON, OFF

Setup.AutoInfo.T.EP ON, OFF

Setup.AutoInfo.T.RC ON, OFF

Setup.AutoInfo.C.B1 ON, OFF

Setup.AutoInfo.C.R1 ON, OFF

Setup.AutoInfo.C.B2 ON, OFF

Setup.AutoInfo.C.R2 ON, OFF

Setup.AutoInfo.PR.B (only at 756)

ON, OFF

Setup.AutoInfo.PR.R (only at 756)

ON, OFF

Setup.AutoInfo.I ON, OFF

Setup.AutoInfo.O ON, OFF

ON means that the Coulometer reports automatically the moment the corre-

sponding change occurs.

.Status:

Global switch for all set AutoInfo.

.P

PowerOn: Simulation of power on (4.2.2.99). Not from mains.

Messages from node .T, Titrator:

.T.R

Ready: Status 'Ready' has been reached.

.T.G

Go: Instrument has been started.

.T.GC

GoCommand: Instrument has received a go command.

.T.S

Stop: Status 'Stop' has been reached.

.T.B

Begin of method.

.T.F

Final: End of determination, the final steps will be carried out.

.T.E

Error. Message together with error number, see page 56ff.

.T.O

Conditioning OK: EP reached.

.T.N

Conditioning Not OK: EP not reached.

.T.Re

Request: In the inquiry of an identification or the sample size after

start of titration.

.T.Si

SiloEmpty: Silo empty, i.e. the last line has been removed from

the silo memory.

.T.M

MeasList: Entry in the measuring point list.

.T.EP

EPList: Entry into EP list.

.T.RC

Results have been recalculated.

Messages from node .C, Comport:

.C.B1

COM1: A report is outputted on COM1. During this time, COM2

will be blocked. COM2 is generally blocked, if COM1 is busy.

.C.R1

COM1 is ready again. (Comes also when you <QUIT> an error.)

.C.B2, .R2 Identical for COM2.

Messages from node .PR, internal printer (only at 756):

Advertising