2 connecting coulometer to dosino, 1 setup with aspiration equipment – Metrohm 756 KF Coulometer User Manual

Page 123

Advertising
background image

6.2 Connecting Coulometer to Dosino

756/831 KF Coulometer, Instructions of Use

115

6.2 Connecting Coulometer to Dosino

Automatic reagent exchange is possible with the Dosino.

The inquiry

reagent change

under <CONFIG>,

>monitoring

must be set to

auto

or "man.".

The key <EXCH> is then used to carry out a reagent exchange.

With reagent change

auto

the reagent change is carried out automatically as soon as the

reagent monitoring has responded.

700

The 2.700.0020 Dosino can be connected directly. If you want to connect a 2.700.0010

Dosino then you require the 6.2134.020 adapter cable.

For aspiration it is an advantage to use the 6.5617.000 aspiration equipment (including

50 ml dosing unit; order Dosino separately).

For aspirating oily samples, where only the sample is to be aspirated and not the whole

reagent, a 20 ml dosing unit or, for very viscous samples, a 10 ml cylinder should be

used; see page 146ff for accessories.

6.2.1 Setup with aspiration equipment

1. Screw 6.1829.010 tube into the threaded opening below the dosing unit. It may be

necessary to cut the tube to the correct length.

2. Screw the dosing unit onto the reagent bottle and insert the reagent bottle from above

at an angle into the bottle holder. Fill the small adsorber tube with molecular sieve and

attach it to the dosing unit. Place the Dosino on the dosing unit.