2.6 Tips for working with water standards

756/831 KF Coulometer, Instructions of Use

9

2.6 Tips for working with water standards

For validation of the instrument, as a fully integrated measuring sys-

tem, commercial, certified water standard solutions with water con-
tents of 1.00

±

0.003 mg/g and/or 0.10

±

0.005 mg/g should be ap-

plied (The 1.0 mg/g Standard is easier to handle and therefore to

prefer).
Recommended initial weight range:

Liquid standard 1.0 mg/g

0.2-2.0 g

Liquid standard 0.1 mg/g

0.5-1.5 g

2.6.1 Recommendations for practice

For validation of the system very accurate handling is needed. To

minimise possible measuring inaccuracies the sample preparation

and handling should run accordingly to the following procedure:

1. Wear gloves (As always in KF Titration).
2. Take a fresh plastic syringe and open it.
3. Take a fresh ampoule of KF standard and shake it for 10

seconds.

4. Open the ampoule and suck 1 ml of the standard into the

syringe

5. Pull the piston of the syringe up to the end and shake the

syringe for a few seconds, so that the inner part of the sy-

ringe is rinsed with standard and gets rid of water

contamination.

6. Splash the used standard into a waste bottle.
7. Repeat the same procedure with another ml of the standard

solution.

8. Suck the whole rest of the standard into your syringe.

Thereafter, verify that there is no more solution in the needle

by sucking a small amount of air into the syringe.

9. Clean the needle by wiping it with a soft tissue. Close the

needle with the corresponding cap.

10. Place the syringe on the balance and press TARA.
11. As soon as the drift at your Coulometer is stable, you can

take the syringe, press <Start> at the Coulometer and in-

ject around 1 ml of the standard. This can be done in two

different ways:

a. The standard is injected without dipping the needle. If

a small drop keeps hanging at the needle, aspirate it

back into the needle, before pulling the needle out of

the septum.

b. The standard is injected directly under the surface of

the KF solution.

Furthermore, make sure that the standard doesn’t splash

on the wall of the vessel or on the electrode.