6.3 Connecting the 768 KF Oven

756/831 KF Coulometer, Instructions for Use

118

6.3.1 Equipping the titration vessel with an oven

The titration vessel is equipped in a similar way to that for aspiration with a Dosino, see

page 116. The gas outlet of the oven is connected to the tip. Take care that connection

between the oven outlet and the titration vessel is short, as otherwise condensed water

may form in the tubing!

If you use the 6.1830.000 heatable outlet tubing then you require 6.1446.170 stopper for

the inlet tip.

If in addition to the gas inlet from the oven you want to use a Dosino for aspiration then

the aspiration tip is inserted on the joint opening and the gas inlet is fitted with A.254.0104

seal which is placed in the screw cap instead of the septum; the screw cap is then

screwed down; see below.

If you use the 6.1830.000 heatable outlet tubing then you should use A.254.0102 seal

(instead of A.254.0104) for the gas inlet.

If you require an additional opening for injections then 6.1465.320 titration vessel is avail-

able; it has two side-mounted screw threaded openings.

Drying tube

Generator electrode

Indicator electrode

Gas inlet

Aspiration tip

A.254.0104 sealing