3 titration vessel setup with ti stand – Metrohm 756 KF Coulometer User Manual

Page 122

Advertising
background image

6.1 Coulometer setup

756/831 KF Coulometer, Instructions for Use

114

6.1.3 Titration vessel setup with Ti Stand

1. Attach titration vessel with holder to the support rod.

2. Place stirring bar in titration vessel.

3. Cut 6.2713.XXX ground joint sleeves to the correct length and use for all the joints of

the inserts

1)

.

4. Insert indicator electrode in the left-hand joint opening, screw on 6.2104.020 electrode

cable and plug it into the "Ind.El" socket of the Coulometer.

5. Insert generator electrode in the central joint opening, screw on 6.2104.120 electrode

cable and plug it into the "Gen.El" socket of the Coulometer.

6. Fill the drying tube with molecular sieve and insert into generator electrode.

7. Place septum in the screw cap and screw this onto the titration vessel. Only tighten it

enough to ensure that it is tight. (The septum should not be deformed!)

8. Insert 6.1439.010 feed/aspiration tube (order separately) in the last joint opening and

connect the aspiration and feed tubing of the Ti-Stand. Close the top of the tube with a

glass stopper.

1)

When cutting the ground joint sleeves take care that no rough edges are formed. The

ground joint sleeves must not project beyond the lower edge of the joint.

If no ground joint sleeves are used then the joints must be greased. In this case the

joints must be checked periodically and re-greased while otherwise problems with

blocked joints could occur.

Drying tube

Generator electrode

Indicator electrode

Septum stopper

Feed/aspiration tube

Inlet

Outlet