5.5 Testing the measuring input

756/831 KF Coulometer, Instructions of Use

111

5.5 Testing the measuring input

With the aid of the "767.0010 Calibrated Reference for mV, pH,

, uS, °C" you can check

the measuring input "Ipol" and the indicator electrode cable.

If a Remote Box is connected:

Deactivate the Remote Box (key <Config>,

>peripheral units

,

Remote Box: off

). Switch

the Coulometer off and screw off the Remote Box. Switch the Coulometer on again (so

that the new configuration will be recognized).

Procedure:

1. Switch off Coulometer.

2. Screw off indicator electrode cable and insert in socket 5 of the 767. The cover re-

mains closed on the 767.

3. Switch on Coulometer and press key <9> at the same time. The display shows:

diagnose press key 0...9

4. Press key <6>. The display shows:

pol/ADC test press 1..3

5. Press key <3>. The display shows:

polarizer test

6. Press <ENTER> on the Coulometer. The display shows:

dummy resistor 10.0 k ?

Press <ENTER> and enter the resistance from the cover of the 767 (

-value 5). The

display shows:

polarizer test *

When the test has been completed the display shows:

polarizer test o.k.

7. Exit the diagnostic program with 3 times <CLEAR>.

8. Make the Coulometer ready for work again:

. Screw the cable back on to the indicator electrode.

. Plug in the cable of the generator electrode.

The measuring input and cable have now been checked.