3.7 Result calculations

756/831 KF Coulometer, Instructions of Use

35


RS1 text RS1


RS1 decimal places 2

RS1 unit: ppm


RS1 limit control: OFF



RS1 low lim. 0.0
RS1 up lim. 0.0

RS1 L13 output: OFF

Text for result output (up to 8 characters)

Text input see page 16.

Number of decimal places for result (0...5)

Selection of result unit (ppm, mg/g, mg/ml, mg, ug,

mg/pc, %, no unit or up to 6 characters).

Limit control for the result (on, off)

The limits are checked each time a result is calculated.

If

on

has been set:

Lower limit (0.0...999 999)

Upper limit (0.0...999 999)

Sets line L13 of the remote socket (OFF, active, pulse)

if the result lies outside the limits.

Enter next formula, e.g. for RS2.

Meaning of the calculation variables CXX:

C00

Sample size, see page 46.

C01...C19

Method-specific operands, see page 36. They are stored with the

method in the method memory.

C21...C23

Sample specific operands, see page 46ff.

C26, 27

Mean values from silo calculations.

C30...C39 Common

variables.

C40

Initial measured value of the sample.

C41

Amount of water at the end of the titration in ug.

C42 Determination

time.

C43

Drift at the start of the titration.

C44 Temperature.

C45

Amount of charge in mA

s.