4.2 Remote control commands

756/831 KF Coulometer, Instructions for Use

86

Info.ActualInfo.Outputs.Clear $G

Status sends the current status of the I/O lines, Change sends the information

regarding whether a change in status of a line has taken place since the last

clearing, Clear clears the change information. For the output, there is a con-

version from binary to decimal, e.g.

| 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 1 | 0 |

Line No.

| 13| 12| 11| 10| 9 | 8 | 7 | 6 | 5 | 4 | 3 | 2 | 1 | 0 |

Output: 2

1

+ 2

3

= "10"

1 means ON or change; 0 means OFF or no change.

The lines are assigned as follows (see also page 131):

Inputs: Outputs:

0

Start (pin 21)

0

Ready (pin 5)

1

Stop (pin 9)

1

Cond. ok (pin 18)

2

Enter (pin 22)

2

Titration (pin 4)

3

pin 10

3

EOD (pin 17)

4

pin 23

4

not used (pin 3)

5

pin 11

5

Error (pin 16)

6

pin 24

6

Activate, line L6 (pin 1)

7

pin 12

7

Pulse for recorder (pin 2)

8

Connected remote box (pin 6)

9

not used (pin 7)

10

not used (pin 8)

11

Change reagent (pin 13)

12

Smpl size out (pin 19)

13

Result out (pin 20)

4.2.2.70. Info.ActualInfo.Assembly.CyclNo read

only

Info.ActualInfo.Assembly.I read

only

Info.ActualInfo.Assembly.Meas read

only

Info.ActualInfo.Assembly.Pot read

only

Info.ActualInfo.Assembly.IPulse read

only

Info.ActualInfo.Assembly.Bur.V read

only

Info.ActualInfo.Assembly.Bur.Clear $G

$Q sends the current values.

.CyclNo: Cycle number of the voltage measurement cycle, e.g. "127". From

the cycle number and the cycle time (see 4.2.2.77), a time frame

can be set up. The cycle number is set to 0 on switching on the

instrument and on every start. It is incremented as long as the in-

strument remains switched on.

.I:

Total charge in mA·s, e.g. "667.48".

.Meas:

Measured value in mV at the indicator electrode, e.g. "104.2".

.Pot:

Voltage at generator electrode.

0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3

means >28 V.

.IPulse:

Current of actual pulse.

1 means 100 mA, 2 means 200 mA, 3 means 400 mA.

.Bur.V:

Dosed volume of connected Dosino in ml, e.g. "5.234".

.Bur.Clear: $G clears the volume counter.