5.1 Troubleshooting

756/831 KF Coulometer, Instructions for Use

104

Problem

Possible causes and remedies

Result too low

Drift correction too large, i.e. the drift was too high at the

start or unstable drift.

Stop drift too high.

Min.rate too low

Sample releases iodine.

Results widely scattered

Inhomogeneous sample? Poor reproducibility of sample

addition?

Drift unstable.

Titration times too long

Wait until drift during conditioning becomes stable.

Amount of water too large, see sample size guidelines on

page 10.

Set stop drift higher.

Set control range smaller, set max.rate higher.