3.4 Configuration, key <CONFIG>

756/831 KF Coulometer, Instructions of Use

21

run number 0




operator level: standard







start delay 0 s



result display: bold






dev.label.



beeps 1






display value: OFF



program 5.756.0010

Current run number for result output (0...9999)

The sample number is set to 0 when the instrument is

switched on and incremented on every determination.

Operating mode (standard, expert)

Determines the number of inquiries which are

accessible. Operation in the standard mode contains

only a few inquiries and is recommended for routine

applications.

Inquiries which are accessible in the standard mode

are highlighted in gray in these Instructions for Use.

Start delay (0...999 999 s)

Delay time after start of methods. Abort start delay time

with <QUIT>.

Type of result display at the end of the determination

(bold, standard)

bold

: the calculated results are displayed in bold

characters.

standard

: displays the whole information, e.g. results,

water, messages etc.

Individual identification of devices (up to 8 ASCII

characters). Is automatically printed in reports.

Number of beeps (1...3, OFF)

when instrument is ready (conditioning OK), end of

titration and Cond.OK, reception of sample data from

the balance and with sample sizes outside the limiting

values.

Display of measured value (ON, OFF)

Display of U-value during conditioning and titration.

Display of program version. At 831: 5.831.0011 ; at 756:

5.756.0012 .


>RS232 settings COM1


baud rate: 9600

data bit: 8

stop bit: 1

parity: none

handshake: HWs

Settings of RS232 interface

see also pages 97ff. Identical for COM2.

Baud rate (300, 600, 1200, 2400, 4800, 9600)

Data bit (7, 8)

Stop bit (1, 2)

Parity (even, odd, none)

Handshake (HWs, SWline, SWchar, none)

see page 97.