2.4 Generator electrode without diaphragm

756/831 KF Coulometer, Instructions of Use

7

2.4 Generator electrode without diaphragm

The 6.0345.100 generator electrode without diaphragm poses no

handling problems and is easy to clean. It only requires one reagent

and is quickly ready for use (no moisture depots in the diaphragm!).

The generator electrode without diaphragm is the best choice for

most applications. It is particularly suitable for use with very polluting

samples.

2.4.1 Reagents

Only use those reagents which are specially intended for use with

generator electrodes without diaphragm; see the reagent manufac-

turer's documentation.

2.4.2 Cleaning

The electrolyte solution can normally be exchanged without any spe-

cial cleaning of the parts being necessary. If cleaning is necessary

then care should be taken that the Pt grid of the generator electrode

is not damaged.

Pollutants containing oil:

Clean with a solvent (e.g. hexane) and then rinse with ethanol.

Salt-like deposits:

Clean with water and then rinse with ethanol.

Dry all parts thoroughly after cleaning. A hot-air blower can be used

for this. If the parts are dried in a drying oven take care that the tem-

perature does not exceed 70°C (plastic components!).