5.2 Error and special messages

756/831 KF Coulometer, Instructions for Use

106

division by zero

The result could not be calculated as a divisor in the formula

was equal to zero.

Exit: Enter appropriate value.

initializing Dosino

When the Coulometer is switched on the connected Dosino is

initialized.

instr.temp.too high

The temperature in the Coulometer is too high (

60 °C).

Exit: Wait until temperature is <60 °C.

manual stop

The determination has been manually stopped.

meas.pt list overflow

Maximum 500 measured points can be stored.

Exit: Select larger time interval.

missing EP

An EP needed for calculation in a formula is missing.

no method

The method required by the sample data from the silo memory

is not available in the method memory.

Exit: <CLEAR>.

no new com.var.

The common variable could not be assigned as the result or

the mean value could not be calculated. The old value remains

in force.

no new mean

No new mean value has been calculated as at least one

quantity stipulated for mean value calculations could not be

calculated.

no oven param.

The oven could not be found at the given COM.

Remedy: connect the oven to the given RS-interface of the

Coulometer or set the following in your method under

<PARAM>, >preselections,

Oven: no

.

no titration data

No curve can be printed as no data are available.

not valid

A value is not available.

overrange

The measuring range of

±

2 V has been exceeded. Overrange

replaces the corresponding measured value.

Exit: Rectify error or <STOP>.

overtitrated

In iodine range. The message can also appear after switching

on. Add methanol. If the message appears again:

- Check whether the cables of the indicator and generator

electrodes have been interchanged.

- Improve stirring.

- Exchange reagent.

Exit: Rectify fault or <STOP>.

result out of limits

The result lies outside the limits which were defined in the

method, see page 35.

Exit: Calculate result again or new start.

sample size out

The sample size is outside the limits which are defined in the

method, see page 29.

Exit: Enter new sample size.

sample unfit

The EP has been "overshot" during the titration. The sample

may release an oxidizing agent or the control parameters have

not been set correctly. The result could be incorrect.

service is due

The service interval has elapsed. Contact Metrohm service so

that the Coulometer can be serviced. This message will ap-