3.10 Data output

756/831 KF Coulometer, Instructions for Use

40

3.10 Data output

3.10.1 Reports for the output at the end of a determination

With <DEF>, the report sequence at the end of the

determination is defined.

The entries are specific to the method and are stored in

the method memory.

Report:

Definition of report blocks to be printed automatically at

the end of the determination.

The display texts of the Coulometer are shown to the

left. The values are the default values. Inquiries which

also appear in the standard operation mode are high-

lighted in gray.

>report
Only at 756:
internal:result;




At both 756 and 831:
COM1:Result;
COM2:Result;

Report sequence

Report sequence for the internal printer (result, water

crv, rate crv, meas crv, comb, mplist, param, calc, scalc

full, scalc srt, ff)

Select a block with keys <

> and <

>. If you

require more than one report block, set a "

;

" as a

separator between the blocks.

Identical for COM1 and COM2.

Meaning of the report blocks:
result

Result report with raw results, calculations and statistics.

water crv

Curve "mass water in ug" vs. time.

rate crv

Curve "rate in ug/min" vs. time.

meas crv

Curve measured voltage vs. time.

comb

Combined curve: mass of water and rate vs. time.

mplist

Measuring point list.

param Parameter

report.

calc

Report with formulas and operands.

scalc full

Full report of silo calculations.

scalc srt

Short report of silo calculations.

ff

Form feed on printer.

Original reports which are put out automatically at the end of the titration can be printed

with recalculated values at any time. Key sequence:

<PRINT><REPORTS><ENTER>.

The target of these reports is as defined in the method.

def

>formula

>silo calculations

>common variables

>report

>mean

DEF

(