2 equipping the titration vessel for aspiration – Metrohm 756 KF Coulometer User Manual

Page 124

Advertising
background image

6.2 Connecting Coulometer to Dosino

756/831 KF Coulometer, Instructions for Use

116

3. Screw 6.1602.105 bottle attachment onto the waste bottle, fill adsorber tube with mo-

lecular sieve and place it in the bottle attachment. Close the larger threaded opening

of the bottle attachment with 6.1446.080 stopper.

4. Place the waste bottle in the bottle holder. Connect the bottle attachment and Port 3 of

the dosing unit with 6.1805.080 tubing.

5. Equip the titration vessel with the aspiration equipment, see below. The aspiration

equipment consists of 6.1543.070 tip, nipple from 6.2730.030 (use E.301.0022 O-ring

instead of the thin O-ring of the nipple) and 6.1446.060 stopper.

6. Connect the aspiration tip and Port 1 of the dosing unit with 6.1805.060 tubing.

6.2.2 Equipping the titration vessel for aspiration

1. Attach titration vessel with holder to the support rod.

2. Place stirring bar in titration vessel.

3. Cut 6.2713.XXX ground joint sleeves to the correct length and use for all the joints of

the inserts

1)

.

4. Insert indicator electrode in the left-hand joint opening, screw on 6.2104.020 elec-

trode cable and plug it into the "Ind.El" socket of the Coulometer.

5. Insert generator electrode in the central joint opening, screw on 6.2104.120 electrode

cable and plug it into the "Gen.El" socket of the Coulometer.

6. Fill the drying tube with molecular sieve and insert into generator electrode.

7. Place septum in the screw cap and screw this onto the titration vessel. Only tighten it

enough to ensure that it is tight. (The septum should not be deformed!)

8. Screw tip with the nipple and O-ring from 6.2730.030 into 6.1446.060 stopper and

insert this into the last joint opening.

9. Connect the tip to Dosino Port 1.

10. Connect Dosino Port 3 to the waste bottle.

1)

When cutting the ground joint sleeves take care that no rough edges are formed. The

ground joint sleeves must not project beyond the lower edge of the joint.

If no ground joint sleeves are used then the joints must be greased. In this case the

joints must be checked periodically and re-greased while otherwise problems with

blocked joints could occur.

Drying tube

Generator electrode

Indicator electrode

Septum stopper

Tip

Nipple from

6.2730.030

6.1446.060

Stopper