4.2 Remote control commands

756/831 KF Coulometer, Instructions of Use

79

Config.Monitoring.Reagent.ReagCap 1...1000...9999, OFF

Config.Monitoring.Reagent.RCounter

0...9999

Config.Monitoring.Reagent.ClearCount $G

Config.Monitoring.Reagent.Drift 0...99,

OFF

Monitoring of reagent live time.

.Determ: Number of determinations.

.DCounter: Counter of determinations already carried out.

.MaxTime: Maximum live time of reagent in days.

.TCounter: Time already elapsed since last reagent change.

.ReagCap: Reagent capacity in mg water.

.RCounter: Counter of reagent capacity.

.ClearCount: Clears all above counters.

.Drift:

Stable drift in ug/min.

4.2.2.32. Config.Monitoring.Change $G,

$S

Config.Monitoring.Change.Status auto,

man.,

OFF

Config.Monitoring.Change.WaitTime

0...999 999

Config.Monitoring.Change.AspVol 0...100...9999

Config.Monitoring.Change.SolventVol 0...100...9999

Config.Monitoring.Change.Rinse

0...9999

Config.Monitoring.Change.NoRinse

1...9

Changing of reagent. With a connected Dosino, the reagent is changed with

&Config.Monitoring.Change $G. The nod &Config.Monitoring.change.Status
has to be

OFF.

Parameters for automatic reagent change:

.WaitTime: Waiting time in s after switching off the stirrer.

.AspVol: Volume in ml of used reagent to be aspirated.

.SolventVol: Volume in ml of new reagent to be added.

.Rinse:

Volume in ml of rinsing reagent.

.NoRinse: Number of rinsing cycles.

4.2.2.33. Config.Monitoring.Validation.Status ON,

OFF

Config.Monitoring.Validation.Interval 1...365...9999

Config.Monitoring.Validation.Counter

0...9999

Config.Monitoring.Validation.ClearCount $G

Monitoring of validation.

.Interval: Time interval in days for validation.

.Counter: Time counter in days since last validation.

.ClearCount: Clears the above counter.

4.2.2.34. Config.Monitoring.Service.Status ON,

OFF

Config.Monitoring.Service.Date XXXX-XX-XX

Monitoring of service interval.

4.2.2.35. Config.Monitoring.DiagRep ON,

OFF

Printing of system test report after each switching on of the Coulometer.

4.2.2.36. Config.PeriphUnit.CharSet1

Epson, Seiko, Citizen, HP, IBM