4.2 Remote control commands

756/831 KF Coulometer, Instructions for Use

74

4.2.2 Description of the remote control commands

4.2.2.1. Mode $G,

$S

Start and stop ($G, $S) of the current method (4.2.2.3)

$G also serves to continue after inquiries of identifications and sample size af-

ter the start (see 4.2.2.15)

4.2.2.2. Mode.Select KFC, KFC-B, BLANK, GLP

Selection of the standard mode.

If a method is selected from the method memory, the node &Mode.Select is

overwritten with the mode of the corresponding user method.

4.2.2.3. Mode.Name read

only

Name of the current method in the working memory. $Q sends 8 ASCII char-

acters. Standard methods carry the name ********. The node can be set

read + write, see 4.2.2.66.


4.2.2.4. Mode.Parameter.CtrlPara.EP

0...50...

±

2000

Setting of the EP in mV.

4.2.2.5. Mode.Parameter.CtrlPara.Control content,

special

Mode.Parameter.CtrlPara.Special.Dyn

1...70...2000

Mode.Parameter.CtrlPara.Special.MaxRate 1.5...2240,

max.

Mode.Parameter.CtrlPara.Special.MinRate 0.3...15...999.9, min.

Mode.Parameter.CtrlPara.Special.Stop.Type drift,

rel.drift

Mode.Parameter.CtrlPara.Special.Stop.Drift 1...5...999

Mode.Parameter.CtrlPara.Special.Stop.Drift 0...5...999

Parameters for setting "special" (4.2.2.5):

.Dyn:

Dynamics in mV.

.MaxRate: Maximum allowed titration rate in ug/min. Max. means maximum

possible rate.

.MinRate: Minimum titration rate in ug/min.

.Type:

Type of stop criterion after drift or switch-off delay time.

.Drift:

Stop drift in ug/min. Applies when "drift" has been selected.

.RelDrift: Relative stop drift in ug/min. Applies when "rel.drift" has been se-

lected. Stops if the drift reaches the current drift at the start of the

method plus the rel.drift value.

4.2.2.6. Mode.Parameter.TitrPara.Direction +,

-, auto

Titration direction.

"auto" means the titration direction is determined automatically by the instru-

ment.

4.2.2.7. Mode.Parameter.TitrPara.Pause

0...999 999

Mode.Parameter.TitrPara.ExtrT

0...999 999

.Pause:

Time in s. During this time, there is no generation of current.