7 appendix, 1 technical specifications – Metrohm 756 KF Coulometer User Manual

Page 135

Advertising
background image

7.1 Technical specifications

756/831 KF Coulometer, Instructions of Use

127

7 Appendix

7.1 Technical specifications

Modes

KFC: Coulometric KF titration

KFC-B: Coulometric KF titration with blank deduction

BLANK: Blank determination

GLP: Validation of the Coulometer

Endpoint indication

Voltametric, AC indication

Ipol: 2, 5, 10, 20 or 30 uA adjustable

Iodine production

Pulse with variable current strength and length

Current at the generator electrode: 100, 200, 400 mA

Titration speed

max. 2.24 mg H

2

O/min

Determination range

10 ug to 200 mg H

2

O

Resolution

0.1 ug H

2

O

Reproducibility

Sample: Reagent manufacturer's standard.
With 10 ug...1000 ug H

2

O:

±

3 ug

With >1000 ug H

2

O:

0.3% or better

Drift compensation

automatic, manual or none

Materials

Housing

Metal, powder coated

Keypad cover

Polycarbonate (PC)

Display

Graphical LCD, 192 x 64 dots

Field: 100 x 37 mm

LED

back-lit

Printer (only at 756)

Built-in thermal printer

Paper width 57 mm

144 pixel or 24 characters per line

Memory

Method storage for approx. 100 methods

Silo memory for sample data and results

Stirrer control

On/off switch manual and coordinated with the titration

process

RS232 interface

2 separate interfaces, each can be configured

for printer, balance or computer connection: Completely

controllable from external control unit

Advertising