Metrohm 756 KF Coulometer User Manual

Page 138

Advertising
background image

7.2 Pin assignment of the "Remote" socket

756/831 KF Coulometer, Instructions for Use

130

Outputs
































pin 17 (Output 3)

pin 3 (Output 4)

pin 16 (Output 5)

pin 1 (Output 6)

pin 2 (Output 7)

pin 6 (Output 8)

pin 7 (Output 9)

pin 8 (Output 10)

pin 13 (Output 11)

pin 19 (Output 12)

pin 20 (Output 13)

End of determination

EOD

not used

Error,

active on errors

Activate pulse, see also page 132.

Pulses for recorder (t

p

=150

µ

s)

Remote box activated

not used

Reagent exchange

Sample size out of limits

Result out of limits (adjustable, see

page 35)

For all outputs:

V

CE0

= 40 V

I

C

= 20 mA

t

Pulse

> 100 ms

Functions see page 131.

Voltage

pin 15

pin 14

pin 25

I

200 mA

0 V: active

5 V: inactive

Ordering numbers for plug:

K.210.9004 (shell) and K.210.002

No liability whatsoever will be accepted for damage caused by im-

proper interconnection of instruments.

0 V

+5 V