Metrohm 756 KF Coulometer User Manual

Page 84

Advertising
background image

4.2 Remote control commands

756/831 KF Coulometer, Instructions for Use

76

.Cond:

Conditioning ON/OFF

.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of

the drift value at start.

.DCor.Value: Drift value for the manual drift correction.

4.2.2.15. Mode.Parameter.Presel.IReq

id1, id1&2, all, OFF

Mode.Parameter.Presel.SReq

value, unit, all, OFF

Mode.Parameter.Presel.ReqTitr

ON, OFF

Automatic inquiries after the start of the determination. From such an inquiry,

the determination continues if the requested entry/entries is/are made, e.g.

&SmplData.OFFSilo.Id1 (see 4.2.2.56) or with &M $G, see 4.2.2.1.

.ReqTitr: Current generation starts during requests (with ON).

4.2.2.16. Mode.Parameter.Presel.SampleUnit

g, ...up to 5 ASCII

Method specific sample unit, i.e. when the method is loaded, the current unit

of the sample size is overwritten by the unit from the method.

4.2.2.17. Mode.Parameter.Presel.LimSmplSize.Status ON,

OFF

Mode.Parameter.Presel.LimSmplSize.LoLim

0.0...999 999

Mode.Parameter.Presel.LimSmplSize.UpLim 0.0...999 999

Limit control for the sample size.

4.2.2.18. Mode.Parameter.Presel.Id1Text

id1/C21, 10 ASCII characters

Mode.Parameter.Presel.Id2Text

id2/C22, 10 ASCII characters

Mode.Parameter.Presel.Id3Text

id3/C23, 10 ASCII characters

Text for sample identifications.

4.2.2.19. Mode.Parameter.Presel.Cell

no diaph., diaphragm

Mode.Parameter.Presel.GenI 100,

200,

400, auto

.Cell:

Type of generator electrode.

.GenI:

Current at the generator electrode in mA. "auto" means that the

current is switched in the course of determinations.

Default: 400 mA for cells without diaphragm, auto for cell with

diaphragm.

4.2.2.20. Mode.Parameter.Presel.Oven COM1,

COM2,

no

If an Oven is connected, its results will be incorporated into the result report of

the Coulometer.

If there is no Oven connected via RS232, this parameter has to be on "no".

4.2.2.21. Mode.Parameter.Presel.ActPuls

first, all, cond., OFF

Output of a pulse on the I/O line "Activate", see page 132.

4.2.2.22. Mode.Def.Formulas.1.Formula

H2O, CXX, RSX, +, -, *, /, (, )

Mode.Def.Formulas.1.TextRS

8 ASCII characters