6.4 Connecting the 774 Oven Sample Processor

756/831 KF Coulometer, Instructions for Use

120

6.4.1 Equipping the titration vessel with the Oven Sample Proc-

essor

Lead the tip of 6.1830.010 heatable outlet tube into the titration vessel with the aid of

6.1446.170 stopper.

Outlet heater

If in addition a Dosino is to be used for aspiration then the aspiration tip is inserted into

the joint opening and the gas inlet is fitted with A.254.0102 seal which is placed in the

screw cap instead of the septum; the screw cap is then screwed down; see illustration on

page 118.

If you require an additional opening for injections then 6.1465.320 titration vessel is avail-

able; it has two side-mounted screw threaded openings.

Stopper with

6.1446.170 O-ring

Generator

electrode

with drying

Connection for

outlet heating

Transfer hose with

outlet heating

6.1830.010

Indicator

electrode

Septum stopper

6.1464.320

Titration vessel

728 Mag-

netic stir-

Fastening screws