7 result calculations – Metrohm 756 KF Coulometer User Manual

Page 42

Advertising
background image

3.7 Result calculations

756/831 KF Coulometer, Instructions for Use

34

3.7 Result calculations

Formula entry, key <DEF>

Key <DEF> contains various inquiries for result

calculations and data output. The data of this key are

method-specific and they are stored in the method

memory together with the method.

Formula (in expert mode only):

Formulas for result calculations.

The display texts of the Coulometer are shown to the

left. The values are the default values.


>formula

RS?




RS1=


RS1=H2O*C01/C00/C02





























Input of formulas

Enter formula number (1...9)

You can calculate up to 9 results per method.

Enter a number 1...9.

Input of formula

Example:

RS1=H2O*C01/C00

Enter formula by means of 3rd functions of keyboard.

Here you will find operands, mathematical operations

and parentheses. Operands require a number as an

identification. You can use the following operands:

H2O

: Amount of water at the EP in ug.

RSX

: Results which have already been calculated with

previous formulas. X = 1...9.

CXX

: Calculation constants. XX = 00...45.

Rules:

Calculation operations are performed in the

algebraic hierarchy: * and / before + and -.

Store formula with <ENTER>.

Calculation quantities and operands can be deleted

with <CLEAR> one by one.

To delete a complete formula press <CLEAR>

repeatedly until only RSX remains in the display.

Confirm with <ENTER>.

If a formula is stored with <ENTER>, result text,