5.2 Error and special messages

756/831 KF Coulometer, Instructions of Use

107

pear each time the Coulometer is switched on.

Exit: New start.

silo empty

The silo memory is switched on but is empty and a titration

has been started.

Corrective action: Fill at least the first 1 silo line before starting

the first titration.

Exit: <CLEAR>.

silo full

The silo memory is full (255 lines).

Exit: <CLEAR>.

stop time reached

The titration has been stopped as the max.titration time has

been reached.

system error 3

The instrument adjustment data have been overwritten.

Exit: <CLEAR>. Default adjustment data are set. The error

message appears each time the instrument is switched on

until it has been readjusted (Metrohm service).

system error 14

No communication between the Coulometer and the con-

nected Remote Box.

Possible causes:

-

The Remote Box was connected when the Coulometer was

running

-

Coulometer has a fault.

-

Remote Box has a fault.

Remedy: Set under <CONFIG>,

>peripheral units

,

Remote

Box: OFF

, switch off Coulometer, take away Remote Box and

switch on Coulometer. Contact Metrohm service.

time-out PC keyboard

A connected PC keyboard has been used to call up an ad-

dress (e.g. <F12>) and the connection has then been inter-

rupted.

Possible causes:

-

Remote Box has a fault.

-

PC keyboard has a fault.

Exit: Correct fault and switch Coulometer off/on.

transmission error

With a Remote Box connected characters are received which

cannot be interpreted.

Possible causes:

-

Wrong key combination has been pressed.

-

Wrong PC keyboard has been selected.

-

The barcode reader supplies garbled characters.

-

The Remote Box has a fault.

Exit: Rectify fault and switch Coulometer off/on.

validate instrument

Validation interval has elapsed.

Exit: <CLEAR> or new start.

work.conditions not ok

During the titration there was too high a resistance at the

generator electrode. The result could be incorrect. Reasons:

- Not enough reagent in the titration vessel.

- Gas bubbles when working with the oven: Set smaller gas

flow at the oven.