4.1 General rules

756/831 KF Coulometer, Instructions of Use

59

E196

Result is out of limits.

Exit: The error message disappears on next start or on recalculation.

E197

Sample size is out of limits.

Exit: The error message disappears on next start or on introduction of new

sample size.

E198

Validation interval is expired.

Exit: The error message disappears on next start or clear counter with

&Config.Monitoring.Validation.Clear $G.

E199

Service date is reached.

Exit: The error message disappears on next start or change date in

&Config.Monitoring.Service.Date.

E203

No Oven parameters: Oven not (correctly) connected.

Exit: The error message disappears on next start. If you don't wish oven

parameters in your report, select &Mode.Parameter.Presel.Oven "no" in your

method(s).

E209

Temperature in the KF Coulometer instrument too high (>60 °C).

Exit: The error message disappears if the Coulometer temperature is below

60°C.

E212

Transmission error from Remote Box. Unknown characters.

Exit: Rectify error and switch Coulometer off and on again.

E213

Time-out error from PC keyboard (Remote Box)

Exit: Rectify error and switch Coulometer off and on again.

E214

Check Remote Box. Remote Box not (properly) connected but activated in

&Config.Periph.RemoteBox.

Exit: Rectify error and switch Coulometer off and on again.