4.2 Remote control commands

756/831 KF Coulometer, Instructions of Use

77

Mode.Def.Formulas.1.Decimal 0...2...5

Mode.Def.Formulas.1.Unit

6 ASCII characters

Mode.Def.Formulas.1.Limits ON,

OFF

Mode.Def.Formulas.1.LoLim

0...

±

999 999

Mode.Def.Formulas.1.UpLim

0...

±

999 999

Mode.Def.Formulas.1.Output active,

pulse,

OFF

Mode.Def.Formulas.2.Formula

etc. up to .9

Entry of formulas. Rules for formula entry, see page 34.

Example: "H2O*C01/C00"

In addition to the formula, a text for result output, the number of decimal

places and a unit for the result output can be selected. "No unit" is selected

with the blank string.

In place of "RSX", a result name may be entered (.TextRS). This name is out-

putted in the result report, scalc full and scalc srt. It is used for the result and

the corresponding mean value.

The limit control for results can also be activated. If a result is out of limit, a

message appears in the result report, E196 is sent, and output line L13 can

be set.

4.2.2.23. Mode.Def.SiloCalc.Assign.C24

RSX, H2O, CXX

Mode.Def.SiloCalc.Assign.C25

RSX, H2O, CXX

Mode.Def.SiloCalc.MatchId

id1, id1&2, all, OFF

.Assign.C2X: Assignment to store results in the silo as C2X.

.MatchId:

Indication which sample identification(s) have to match so that

the results can be combined.

4.2.2.24. Mode.Def.ComVar.C30

RSX, MNX, H2O, CXX

Mode.Def.ComVar.C31

etc., up to .C39

Assignment of common variables.

The values of the common variables are to be found in &Config.ComVar. They

can be viewed and entered there, see 4.2.2.54.

4.2.2.25. Mode.Def.Report.Internal (only 756) result, water crv, rate crv,

meas crv, comb, mplist, param, calc, scalc full, scalc srt, ff

Mode.Def.Report.Assign1 ditto

Mode.Def.Report.Assign2

ditto

Definition of the report sequence, which is outputted automatically at the end

of the determination. Entries of more than one block have to be separated with

";".

.Internal: Internal printer of the Coulometer. (only 756)

.Assign1: Output to COM1 of the Coulometer.

.Assign2: Output to COM2 of the Coulometer.

4.2.2.26. Mode.Def.Mean.1.Assign RS1, RSX, H2O, CXX

Mode.Def.Mean.2.Assign

etc., up to .9