5.4 Initialize KF Coulometer

756/831 KF Coulometer, Instructions for Use

110

5.4 Initialize KF Coulometer

In rare cases the RAM of the Coulometer may need to be initialized. This causes the dele-

tion of all methods, silo data and results. Whenever possible you should first make a

method backup with the aid of a PC and the 6.6008.200 or 6.6008.500 Vesuv 3 Software

and print out your configuration data (<PRINT><CONFIG><ENTER>).

Initialize RAM
1. Switch off Coulometer
2. Switch on Coulometer and press key <9> at the same time. The display shows:

diagnose press key 0...9

3. Press key <8>. The display shows:

RAM init.

4. Press key <ENTER>. Initialization will be carried out. The display then shows:

RAM init. passed

5. Exit the display with <CLEAR>.
6. Re-enter your configuration data and load your methods into the instrument again.