2.3 Your first determination

756/831 KF Coulometer, Instructions for Use

6

2.3 Your first determination



KFC ********


KFC wait
drift
53 ug/min


KFC ready
drift

4.3 ug/min



smpl size 1.0 g








KFC ready
drift

5.3 ug/min

content 38.5 ppm

The titration vessel has been prepared (see page 5) and the

Coulometer is switched on. In the display appears

Press the <START> key.

Pre-conditioning begins, i.e. the titration vessel is dried. The

"COND" LED blinks. The arrow in the drift display shows the

drift tendency (falling, rising, stable).

When the titration vessel is dry an acoustic signal is heard

and the "COND" LED shows a steady light.

Press <START> and inject the first sample.

Enter the sample size and confirm it with <ENTER>.

During the titration you will see the curve µg H

2

O against

time. To the left of the curve the following measurements are

displayed:

H

2

O in µg

Rate in µg/min

Time in s

After the titration the result is displayed and printed out by the

internal printer (with the 831, a printer needs to be installed;

see page 121). The titration vessel is continuously kept dry

and the current drift is displayed.

If you want to determine further samples press <START>

again and inject the next sample...

Advertising