4.2 Remote control commands

756/831 KF Coulometer, Instructions of Use

75

.ExtrT:

Extraction time in s. During this time controlling occurs but the ti-

tration will not be stopped.

4.2.2.8. Mode.Parameter.TitrPara.StartDrift

1...20..999

StartDrift in ug/min. Drift for "conditioning ok" and start of titration possible.

4.2.2.9. Mode.Parameter.TitrPara.Ipol 2,

5,

10, 20, 30

Mode.Parameter.TitrPara.PolElectrTest

ON, OFF

.Ipol:

Selection of polarization current.

If the test for polarized electrodes is switched on, it is performed on change-

over from the inactive state to an active state (titration or conditioning).

4.2.2.10. Mode.Parameter.TitrPara.Temp -170.0...25.0...500.0

Titration temperature in °C.

4.2.2.11. Mode.Parameter.TitrPara.TDelta

1...2...999 999

Time interval in s for the entry of a measurement point in the list of measured

points.

4.2.2.12. Mode.Parameter.TitrPara.TMax

1...999 999, OFF

Maximum titration time in s. After this time, the titration will be stopped.

4.2.2.13. Mode.Parameter.Statistics.Status ON, OFF

Mode.Parameter.Statistics.MeanN

2...20

Mode.Parameter.Statistics.ResTab.Selected original, delete n,

delete

all

Mode.Parameter.Statistics.ResTab.DelN

1...20

Entries for the statistics calculations.

.Status:

On/off switching. Requirement for statistics calculations is a valid

assignment, see 4.2.2.26.

.MeanN: Number of individual results for statistics calculations.

.ResTab.Select: Selection of the table for the statistics calculations.

original: Original table. The original table is (again) set up, i.e. any

individual results which have been deleted are reincorporated

in the statistics calculations.

delete n: Single result lines are removed from the statistics calcu-

lation. All results of the corresponding line in the statistics

table are deleted. Specification of the line number in

.ResTab.DelN.

delete all: Clear entire statistics table. The results can not be reac-

tivated.

.ResTab.DelN: Specification of the line number to be deleted.

4.2.2.14. Mode.Parameter.Presel.Cond

ON, OFF

Mode.Parameter.Presel.DCor.Type

auto, man., OFF

Mode.Parameter.Presel.DCor.Value 0.0...99.9