5.3 Problem with an external printer

756/831 KF Coulometer, Instructions of Use

109

5.3 Problem with an external printer

Problem

Questions for remedial action

No characters can be

received on a connected

printer.

Are the instruments switched on and cables plugged

in correctly?

Is the printer set to ”on-line”?

Are baud rate, data bit and parity the same on both

instruments?

Is the handshake set properly?

If everything seems to be OK, try to print a report with

the key sequence <PRINT><SMPL DATA><ENTER>.

If this report is printed out correctly, check if reports are

defined in key <DEF>.

No data transmission and the

display of the Coulometer

shows an error message.

error 42: Transmission error. Is the printer set to "on-

line"? Is the connection cable properly wired?

The received characters are

garbled.

Are the RS settings the same on both devices?

Has the correct printer been selected?

Data transfer has been interrupted on the hardware

side during the printout of a curve. Re-establish

connections and switch printer off/on.

Wrong line spacing.

The printer does not emulate completely the preset

mode. Usually these problems arise with the IBM mode.

Set the printer to a different mode (e.g. Epson).

Printout of titration curve is

not OK. Other reports are

printed OK.

Handshake is necessary for the printout of curves.

Is your cable correctly wired? (The DTR of the printer

has to be connected to the CTS of the Coulometer.)

Set "HWs" for the handshake of the Coulometer.

Configure the printer such that its DTR is set

(possibly with DIP switches).