3.7 Result calculations

756/831 KF Coulometer, Instructions for Use

36

Input of method-specific operands C01...C19, key <C-FMLA>

With <C-FMLA> the operands C01...C19 can be

entered. For the calculation the operands which were

introduced in the formula are used.

The inputs are method-specific and are stored in the

method memory.

The calculation report can be printed with the key sequence

<PRINT><

/

> (press keys repeatedly until "calc" appears in the display) <ENTER>

Operands C01 and C02
The following table gives the values for the operands C01 and C02 for the standard for-

mulas in the modes KFC and KFC-B depending on the unit in which you want the result to

be expressed and the unit in which you want to enter the sample size:


Result

in

Sample

size in

C01 C02 Result

in

Sample

size in

C01 C02

ppm

%

mg/g

g

1

1

1

1

10 000

1 000

mg/ml

ml

1

1 000

ppm

%

mg/g

mg

1 000

1

1

1

10

1

mg/ml

ul

1

1

C-FMLA

C

1

Advertising