Metrohm 756 KF Coulometer User Manual

Page 99

Advertising
background image

4.2 Remote control commands

756/831 KF Coulometer, Instructions of Use

91

4.2.2.89. Setup.Lock.Keyboard ON, OFF

Setup.Lock.Config ON, OFF

Setup.Lock.Parameter ON, OFF

Setup.Lock.SmplData ON, OFF

Setup.Lock.UserMeth.Recall ON, OFF

Setup.Lock.UserMeth.Store ON, OFF

Setup.Lock.UserMeth.Delete ON, OFF

Setup.Lock.Exchange ON, OFF

Setup.Lock.Display ON, OFF

ON means disable the corresponding function:

.Keyboard: Disable all keys of the Coulometer

.Config:

Disable the <CONFIG> key

.Parameter: Disable the <PARAM> key

.SmplData:

Disable the <SMPL DATA> key

.UserMeth.Recall: Disable "recall" in <USER METH> key

.UserMeth.Store: Disable "store" in < USER METH > key

.UserMeth.Delete: Disable "delete" in < USER METH > key

.Exchange:

Disable the <EXCH> key

.Display:

Disable the display, i.e. it will not be written to by the device

program of the Coulometer and can be operated from the

computer.

4.2.2.90. Setup.Mode.StartWait ON, OFF

Setup.Mode.FinWait ON, OFF

Holding points in the method sequence. If they are "ON", the sequence stops

until "OFF" is sent. Switching the instrument on sets both nodes to OFF:

.StartWait: Holding point right after starting a method (holding point after

AutoInfo !".T.GC").

.FinWait: Holding point at the end a method (holding point after AutoInfo

!".T.F").

4.2.2.91. Setup.SendMeas.SendStatus ON, OFF

Setup.SendMeas.Interval 0.4...4...16200, MPList

.SendStatus: ON means the automatic transmission of measured values (see

4.2.2.94 and 4.2.2.95) in the inputted interval is active.

.Interval: Time interval (in s) for the automatic transmission of associated

measured values defined under points 4.2.2.95 and 4.2.2.96. The

inputted value is rounded off to a multiple of 0.4. The smallest

possible time interval depends on the number of measured values

which have to be sent, on the baud rate, on the load on the inter-

face and on the type of device connection. With "MPList" the

measured values are sent at the time of their entry into the meas-

ured point list.

The automatic transmission is switched on/off with 'SendStatus'.

4.2.2.92. Setup.SendMeas.Select Assembly,

Titrator

Selection of the unit of which the measured values should be sent (4.2.2.95

or 4.2.2.96).