3.4 Configuration, key <CONFIG>

756/831 KF Coulometer, Instructions for Use

18



number of determ. 1






determ.counter 0



reagent lifetime 7 d


time counter 0 d



reagent capacity 1000 mg






capacity counter 0 mg



drift OFF ug/min







reagent change: OFF









waiting time 0 s






If

on

has been set:

Monitoring according to the number of determinations

carried out (1...999, OFF)

The number of determinations which can be carried out

depends on the type of sample (very polluting,

lowering the conductivity) and on the amount of

sample which is to be injected.

OFF

means that monitoring is not active.

Determination counter (0...999)

Counts the number of determinations carried out since

the last time the counters were reset to zero.

Monitoring according to the lifetime of the reagent

(1...9999 d, OFF)

OFF

means that monitoring is not active.

Time counter (0...9999 d)

Counts the number of days since the last time the

counters were reset to zero.

Monitoring the reagent capacity (1...9999 mg, OFF)

With the generator electrode without diaphragm and a

filling volume of 100 ml the capacity is 1000 mg water.

For the generator electrode with diaphragm the

capacity of the katholyte is 300 mg (with 5 ml filling

volume).

OFF

means that monitoring is not active.

Counting the capacity (0...9999 mg)

Adds the weight of water since the last time the

counters were reset to zero.

Monitoring of drift (0...99 ug/min, OFF)

If the current drift value is stable for 2 minutes and

above the set value for drift monitoring (but not

max.=2240 ug/min), the message "change reagent"

appears.

OFF

means that monitoring is not active.

Reagent exchange (auto, man., OFF)

auto

: the reagent is automatically exchanged by the

connected Dosino when the reagent monitoring re-

sponds (see above). The reagent can also be ex-

changed manually at any time with <EXCH>.

man.

: the reagent can be exchanged with <EXCH>.

The reagent exchange procedure is described on page

255.

OFF

: the key <EXCH> is not active.

If "auto" or "man." has been set:

Waiting time before aspiration (0... 999 999 s)

E.g. the waiting time can be used in order to wait for

the phase separation between sample and reagent

when the sample is to be aspirated off.