7 sample addition, 1 sample size, 2 liquid samples – Metrohm 756 KF Coulometer User Manual

Page 18

Advertising
background image

2.7 Sample addition

756/831 KF Coulometer, Instructions for Use

10

12. Close the syringe and put it back on the balance.
13. Read the indicated value off the balance and feed it at your

Coulometer as sample size.

14. As soon as the determination has finished and the titration

cell is conditioned again, you can start with the next deter-

mination.

2.7 Sample addition

This section contains some information about sample addition. A de-

tailed description of this topic is not possible here. You can find fur-

ther information in the reagent manufacturer's documentation and in

Metrohm Application Bulletins.

Metrohm Application Bulletins:

No. 142: Karl Fischer water determination in gaseous samples

No. 145: Determination of small amounts of water in plastics

No. 209: Water determination in insulating oils, hydrocarbons and

their products

No. 273: Validation of KF Coulometers according to GLP/ISO 9001.

2.7.1 Sample size

The sample size should be small so that as many samples as possi-

ble can be titrated in the same electrolyte solution and the titration

time kept short. However, take care that the sample contains at least

50 µg H

2

O. The following table provides guidelines for the sample

weight.

Content of sample

Sample weight

H

2

O to be determined

100000 ppm = 10 %

10000 ppm = 1 %

1000 ppm = 0.1 %

100 ppm = 0.01 %

10 ppm = 0.001 %

50 mg

10 mg... 100 mg

100 mg...

1 g

1

g

5

g

5000 µg

100 µg...1000 µg

100 µg...1000 µg

100 µg

50

µg

2.7.2 Liquid samples

Liquid samples are added with the aid of a syringe. Either a syringe

with a long needle is used with the needle being immersed beneath

the surface of the reagent during injection or a short needle is used

with the last drop being sucked back into the needle.

The best way of determining the actual sample weight is by weighing

the syringe before and after injection.

Volatile or low-viscosity samples should be refrigerated before

that sample is taken in order to prevent handling losses. In contrast,

the syringe itself should not be directly refrigerated as this could

cause the formation of condensate. For the same reason aspirating

air into a syringe which has been cooled by taking up a refrigerated

sample should be avoided.

Advertising