1 reagent exchange procedure with dosino – Metrohm 756 KF Coulometer User Manual

Page 33

Advertising
background image

3.4 Configuration, key <CONFIG>

756/831 KF Coulometer, Instructions of Use

25

3.4.1 Reagent exchange procedure with Dosino

Reagent exchange is automatic (if a reagent monitoring

responds) or is triggered with <EXCH>. During the ex-

change

changing reagent

appears in the display.

Current production and stirrer are switched off.

The waiting time is allowed to elapse. In this time it is pos-

sible to wait for the separation of e.g. a 2-phase mixture.

In this way it is possible to aspirate only 1 phase (e.g. oil

samples).

The given volume is aspirated. A volume slightly larger

than that which is actually to be aspirated should be en-

tered if you want to empty the titration vessel completely.

Rinsing the titration vessel. The rinsing volume is added,

the stirrer switched on for 10 s, and then the rinsing vol-

ume (+3 ml) is aspirated off again. This process is re-

peated for each rinsing cycle.

Normally rinsing is not necessary.

The reagent volume is added and the tubing emptied.

The stirrer is switched on again and the titration vessel is

conditioned.

Basically the instrument is in the same status after the re-

agent exchange as it was before.

<EXCH>

or

Automatic exchange

Conditioning off

Stirrer off

(Waiting time)

(Rinsing volume)

(Rinsing cycles)

Reag.volume

Aspiration volume

Stirrer on

Conditioning on