4.1 General rules

756/831 KF Coulometer, Instructions for Use

58

RS send errors:

E42

CTS=OFF No proper handshake for more than 1 s.

Exit: <QUIT> Is the receiver switched on and ready to receive?

E43

The transmission of the KF Coulometer has been interrupted with

XOFF for at least 6 s.

Exit: Send XON or <QUIT>

E45

The receive buffer of the KF Coulometer contains an incomplete

command (L

F

missing). Sending from the KF Coulometer is

therefore blocked.

Exit: Send L

F

or <QUIT>.


E120

Overrange of the measured value.

Exit: Correct error or &m $S.

E121

Measuring point list overflow (more than 500 measuring points).

Exit: The error message disappears on next start.

E123

Missing EP for calculation.

Exit: The error message disappears on next start or on

recalculation.

E127

Stop time reached.

Exit: The error message disappears on next start.

E128

No new mean.

Exit: The error message disappears on next start or on

recalculation.

E129

No new common variable, old value remains.

Exit: The error message disappears on next start or on

recalculation.

E132

Silo empty and it has been started with open silo or empty silo has

been opened.

Exit: Send a silo entry.

E133

Silo full.

Exit: Send new command.

E134

No method. A method, which is required from the silo memory, does not

exist.

Exit: The error message disappears on next start.

E137

XXX Bytes are missing so that the method, the silo line could not be stored.

Exit: Send new command.

E155

No new silo result (C24 or C25).

Exit: The error message disappears on next start or on recalculation.

E176

The function &Assembly.Buret.Prep or &Assembly.Buret.Empty was

interrupted manually.

Exit: The error message disappears on next start.

E190

Overtitrated. The KF Coulometer is in the Iodine range.

Exit: The error message disappears when the Coulometer is again in the water

range or on next start.

E192

Check generator electrode: Not sufficient solvent in titration vessel or you are

working with fixed generator current or generator electrode defective. The

results of a determination may be erroneous and in the report you will find the

message " work.conditions not ok".

Exit: Rectify error.

E194

Sample unfit. Sample releases oxidative agents during titration.

Exit: Rectify error or &m $S.