6 parameters, key <param – Metrohm 756 KF Coulometer User Manual

Page 35

Advertising
background image

3.6 Parameters, key <PARAM>

756/831 KF Coulometer, Instructions of Use

27

3.6 Parameters, key <PARAM>

The key <PARAM> is used to enter mode-specific

parameters. Values marked with

cond.

are accessible

during conditioning, while

**titr

. means that these

values can also be altered during the titration. In this

case they will influence the run being carried out. All

other values can only be altered in the inactive basic

status.

Two different operating modes are available: standard

mode and expert mode. Inquiries which appear in the

standard mode are highlighted in gray.

The Coulometer displays are shown below at the left-

hand side. The values are the default values.









Control parameters (only in expert mode):

Control parameters for EP.

Titration parameters

Influence the course of the titration.

Statistics:

Mean values and standard deviations of the calculated

results, see page 37.

Preselections:

Selection of various auxiliaries: Automatic inquiries after

the start, etc.


>control parameters

EP at U 50 mV



dynamics 70 mV

**titr.



max.rate max. ug/min

**titr.




min.rate 15 ug/min

**titr.








Control parameters

Endpoint (0.. ±2000 mV)

The standard value should be suitable for most

applications.

Control range 0...2000 mV):

Input as distance to endpoint. Outside the control

range iodine will be produced continuously.

Maximum rate (1.5...2240 ug/min, max.)

<CLEAR> sets

max

.

This parameter primarily determines the rate outside

the control range.

Minimum rate (0.3...999.9 ug/min, min.)

<CLEAR> sets

min

. = 0.28 ug/min.

This parameter determines primarily the rate at the

beginning and at the end of the titration.

parameters

>control parameters

>titration parameters

>statistics

>preselections

PARAM