3.4 Configuration, key <CONFIG>

756/831 KF Coulometer, Instructions for Use

20


man.reports to:
int. (only 756)
COM1
COM2


balance: Sartorius









stirrer control: ON





remote box: OFF



keyboard: US




barcode: input

Target for the output of manually triggered reports (1, 2,

1&2 and only at 756: int., 1&int., 2&int, all)

Manually triggered reports e.g. with <PRINT> .... .

Exception <PRINT><REPORTS>: These reports are

outputted at the target as defined in the method.

Selection of balance (Sartorius, Mettler, Mettler AT,

AND, Precisa)

Sartorius

: Models MP8, MC1

Mettler

:

Models AM, PM and balances with 011,

012, and 016 interfaces

Mettler AT

: Model AT

AND

:

Models ER-60, 120, 180, 182, FR-200, 300

and FX-200, 300, 320

Precisa

:

Models with RS232C interface

Automatic switching ON/OFF of the stirrer in the titration

sequence (ON, OFF)

If stirrer control is

ON

, the stirrer will be switched

automatically. For stirrer control the red switch on the

stirrer unit must be ON.

Connection of a remote box (on ,OFF)

To the remote socket for PC keyboard and barcode

reader, see page 124.

If

on

has been set:

Type of PC keyboard (US, German, French, Spanish,

Swiss.)

The PC keyboard is used as an input aid, see page

125.

Target for barcode reader (input, method, id1, id2, id3,

smpl size)

The barcode reader is used as an input aid, see page

124.

Input

:

The barcode string goes to the entry field in

which the cursor is currently located.

Method

:

The barcode string goes to the entry field

"Methods" in the silo memory.

Id1

:

The barcode string goes to the entry field

"Id1". (Similar for Id2 and Id3.)

Smpl size

: The barcode string goes to the entry field

"smpl size".


> auxiliaries

dialog: english


date 1998-04-23


time 08:13

Various auxiliary settings

Selection of dialog language (english, deutsch,

francais, español, italiano, portugese, svenska)

Current date (YYYY-MM-DD)

Format: year-month-day, entry with leading zeros.

Current time (HH-MM)

Format: hours-minutes, entry with leading zeros.