2.7 Sample addition

756/831 KF Coulometer, Instructions of Use

11

Highly viscous samples can be warmed to lower their viscosity;

the syringe must also be warmed. The same goal can also be

reached by dilution with a suitable solvent. In this case the water

content of the solvent must be determined and deducted as a blank

value correction.

Pastes, greases can be placed in the measuring cell by using a

syringe without a needle. The joint opening can be used for this

purpose. If aspiration is additionally required the opening with the

septum stopper can be used.

The best way of determining the actual sample weight is by weighing

the syringe before and after injection.

With samples containing a lot of water care must be taken that

the needle is not introduced into the measuring cell through the

septum before <START> has been pressed as otherwise the drift

and therefore the result of the analysis could be falsified.

With samples containing only a trace of water the syringe must

be thoroughly dried beforehand. If possible the syringe should be

rinsed with the sample solution by taking up the sample solution

several times and then discarding it.

2.7.3 Solid samples

Whenever possible solid samples should be extracted or dissolved

in a suitable solvent and the resulting solution injected; a blank value

correction should be made for the solvent.

If no suitable solvent can be found for a solid sample or if the

sample reacts with the Karl Fischer solution the drying oven should

be used.

If solid samples have to be placed in the measuring cell directly then

the generator electrode without diaphragm should be used. The

sample can be added through either the joint opening or through the

opening at the side. Take care that:

The sample releases its moisture completely

No side reaction occurs with the Karl Fischer solution

The surface of the electrodes is not covered by the sample

substance (incomplete KF reaction!)

The Pt grid of the generator electrode is not damaged

The Pt wires of the indicator electrode are not damaged