3.4 Configuration, key <CONFIG>

756/831 KF Coulometer, Instructions of Use

17

3.4 Configuration, key <CONFIG>

The key <CONFIG> is used for the entry of instru-

ment-specific data. The set values apply for all modes.

All entries are only possible in the inactive basic status

of the Coulometer.

Two different operating modes are available: standard

mode and expert mode. Inquiries which appear in the

standard mode are highlighted in gray.

Monitoring functions (only in expert mode):

Monitoring the reagent, validation interval, service in-

terval and printout of diagnostic reports.

Peripheral units (only in expert mode):

Selection of printer, balance, PC keyboard, barcode

reader, stirrer control and selection of the COMs for

manual report output.

Auxiliaries:

e.g. selection of operating mode, setting dialog lan-

guage, date, time.

Settings for RS-COM1 and 2 (only in expert mode):

RS parameters for the interfaces.

Report (only in expert mode):

Configuration of the report.

Common Variable (only in expert mode):

Values of the common variables.

The display texts of the Coulometer are shown to the

left. The values are the default values.


>monitoring

reagent: OFF


















Monitoring functions

Monitoring the reagent (ON, OFF)

Monitoring is carried out at the end of the titrations and

when the Coulometer is switched on. If a monitoring

function responds the message "change reagent"

appears. The message vanishes when the reagent is

changed automatically or with <EXCH>. The message

can also be cleared with <CLEAR>. At the same time

all counters are reset to zero.

For generator electrodes with diaphragms the katholyte

normally needs to be changed more frequently than

the anolyte.

CONFIG

configuration

>monitoring

>peripheral units

>auxiliaries

>RS232 settings COM1

>RS232 settings COM2

>report

>common variables