7.4 User methods

756/831 KF Coulometer, Instructions of Use

139

Blank value method, parameters:

'pa

756 KF Coulometer 5.756.0010

date 1997-11-19 time 17:56 0

BLANK 774-Blk

parameters

>control parameters

EP at U 50 mV

dynamics 70 mV

max.rate max. ug/min

min.rate 15 ug/min

stop crit: rel.drift

rel.drift 5 ug/min

>titration parameters

pause 0 s

extr.time 180 s

start drift 10 ug/min

I(pol): 10 uA

electrode test: ON

temperature 25.0 °C

time interval 2 s

max.titr.time OFF s

>statistics

status: ON

mean n= 3

res.tab: original

>preselections

drift corr: auto

req.ident: OFF

req.smpl size: OFF

smpl unit: g

limit smpl size: OFF

text id1 id1 or C21

text id2 id2 or C22

text id3 id3 or C23

cell: no diaph.

generator I: 400 mA

oven: COM1

activate pulse: OFF

------------