8

Safety warnings and cautions

Welch Allyn

WARNING Devices connected to the central station must be certified for overall
system compliance according to the IEC 60601-1 safety standard. The
interconnection of any device with the central station must comply with IEC
60601-1-1. If in doubt about network connectors or devices, please consult your
facility’s Biomedical Engineering department or Welch Allyn Technical Support

Caution The system many not function properly if components have been
dropped or damaged. Protect system components from severe impact or shock.
Do not us the system if you notice any signs of damage.

Caution Do not operate the system in the presence of magnetic resonance
imaging (MRI) or hyperbaric chambers.

Caution No component-level repair of circuit boards and subassemblies is
supported. Use only the support procedures documented in this manual.

Caution Changes to the network interfaces on a live network will cause
outages, and should be planned with the facility. Additional changes may also be
required on each central station if the IP address of the server that supports
them is changed. Refer to the Connex CS Central Station Install Guide and/or
the Connex CS Server Install Guide for additional information.

Caution Customer networks used to support continuous monitoring functions
as part of Connex CS shall be configuration and maintained by information
technology professionals.

Caution Connex CS shall be serviced by Welch Allyn trained service personnel.

Caution Connex CS components shall be replaceable by Welch Allyn service
personnel, authorized customer biomedical engineers, or authorized information
technology professionals.

Caution Do not enable automatic updates on the central station virus protection
software. Automatic updates can impact product performance.