Install a hard disk drive – Welch Allyn Connex CS Central Station & Server - User Manual User Manual

Page 154

Advertising
background image

148

Installing a Replacement Hard Drive

Welch Allyn

Install a hard disk drive

1.

With the replacement drive circuit-side down, position the connector end of the drive
so it is facing the read of the drive carrier.

2.

Align the holes in the drive to the holes in the carrier.

3.

Attach the drive to the carrier with the screws previously used. See letter “D” in

“Figure C-8: Install replacement drive into carrier”

.

Figure C-8: Install replacement drive into carrier

4.

With the carrier’s black lever in the fully open position, slide the drive assembly into
the server. See letter “E” in

“Figure C-9: Install replacement drive assembly into

server”

.

Figure C-9: Install replacement drive assembly into server

5.

Do not push on the black drive carrier lever until the lever begins to close by itself.

6.

When the black drive carrier begins to close by itself, push on it to lock the drive
assembly into place. See letter “F” in

“Figure C-9: Install replacement drive assembly

into server”

.

7.

If the server is currently running, confirm that the green disk activity indicator is
visible on the left side of the carrier. A red indicator may also be present, indicating
that the new disk drive is beginning to synchronize with the existing disk drive.