Connex CS Admin Guide

Localize Station settings

77

Configure alarm hold off

By default, all alarms are set to zero seconds of hold off, meaning they will be displayed
ana annunciated by alarm sound immediately upon occurrence. Hold off settings are
configurable on a per level basis from zero to twenty seconds.

To configure Alarm hold off:

1.

From the Settings > Advanced settings (if required) > Stations Settings tab, click
on Alarm hold off. The alarm hold off configuration window appears.

2.

Select the Edit to make changes.

3.

Click and drag the desired slider for each alarm type Lethal, High, Medium, and Low
as desired. An example is shown if

“Figure 12-5: Alarm holdoff configuration window

example”

on page 77 below.

4.

Click on the check box to enable Patient rest mode allowed. Refer to the Connex CS
Customer Project Req. Form, Appendix C
for input on this clinical setting.

5.

Click Save when finished, and click Back to return to the main Station Settings tab.

Figure 12-5: Alarm holdoff configuration window example

6.

When the enabled, Patient rest mode introduces a layer after the Settings window but
before the requirement enter a User ID and Password. This allows clinical users to
quickly manage Patient rest mode (on / off) without having to login with their user
credentials. See

“Figure12-6: Settings > Patient rest mode control screen composite

example”

on page 78 for a sample of controls available.

7.

The “Advanced settings” button takes users to the login control window to manage
Connex CS settings on the central station.