Late adt settings – Welch Allyn Connex CS Central Station & Server - User Manual User Manual

Page 91

Advertising
background image

Connex CS Admin Guide

Localize Station settings

85

Late ADT settings

1.

Settings starting with “LATEADT” control automatic discharge behavior for patients
which have been under Continuous Monitoring or Spot Vitals taken, but the ADT
message arrived late or after the fact. Some addition settings also control behavior
related to late ADT settings.

Figure 12-15: Late ADT auto discharge related settings

2.

The table below describes the function of each setting.

Table 12-3: LATEADT setting related functions and default values.

Setting Label

Function

Default value

ENABLED

Turns Auto Discharge functions on/off for patients
which have been under Continuous Monitoring or
Spot Vitals taken, but the ADT message arrived late
or after the fact.

TRUE (on)

EXECUTIONFREQUENCY.MINUTES

Controls how often the process runs, in MINUTES

2

EXECUTIONSEQUENCE

Controls the run order between Late ADT Matching,
Confirmed Auto Discharger, and ADT Auto Discharger
processes.

1 (first)

INCLUDENULLADMITDATES

Determines a period of time in MINUTES until
automatic discharge of a patient for which no activity
has taken place for the patient including ADT
messages, Continuous Monitoring, Spot Vitals taken,
edits to demographics or visit data.

FALSE (no)

PATIENTMATCHINGRULE

Determines which matching rule to apply.

1 - if a single match is found, it is used.

2 - if multiple matches are found, try to find a
covering visit: use the ADT visit whose admit
date is closest to the confirmed visit admit date
and whose admit and discharge encompass the
confirmed visit admit and discharge.

3 - If multiple matches are found but a covering
visit cannot be identified, use the best fit visit:
use the ADT visit whose admit date is closest to
the confirmed visits admit date

1

PATIENTOPENVISIT.TIMEELAPSED

INMINUTES

Determines a period of time in MINUTES for how far
back in history search for a patient match.

1440 (24 hours)

Caution Do not change settings for EXECUTION SEQUENCE unless directed
by Welch Allyn Engineering.

Advertising