Configure units of measure – Welch Allyn Connex CS Central Station & Server - User Manual User Manual

Page 86

Advertising
background image

80

Localize Station settings

Welch Allyn

Configure units of measure

Changes to the Connex CS units of measure may be required as driven by customer
preferences.

To change units of measure:

1.

From the Settings > Advanced settings (if required) > Admin tools > Settings
menu, click on Stations. The Station Settings window appears.

2.

Above the Station Settings window header, click on Settings for: CONNEX. A menu
tree appears. An example is shown below in

“Figure 12-9: Station settings menu tree

and hierarchy”

.

Figure 12-9: Station settings menu tree and hierarchy

3.

Click on STATIONS to make change that affect all Central Stations on the network.
The station settings menu changes and the units of measure choice appears.

Caution Although the menu tree allows for making changes to a specific
Central Station, it is generally not recommended to have Central Stations and
devices configured with different units of measure. In typically practice, units of
measure are usually set at the facility level and applied to all systems and
devices.

4.

Click on Units of Measure. The Stations Settings menu tree window closes, and the
hierarchy label appears as Settings for: CONNEX/STATIONS.

5.

Click on Units of measure again. The Units of measure window appears. Click Edit to
view options and make changes. See an example in

“Figure 12-10: Units of measure

window.”

on page 81.

Caution Making changes to Units of measure on a live system or network will
require a reboot of the Central Station / ALL Central Stations and all attached
devices to ensure that all affected systems and devices are operating with the
new settings. Plan your work and inform the clinical staff accordingly.