KEYENCE DV-90 User Manual

Page 31

Advertising
background image

2-7

2

Initial Setup

of the Cod

e

Reade

r

E DV-90-IM

5

Configure the PC communication conditions on the setup software.

„ BL SetUp Software

Configure “PC Port Setting” as follows:
• When connected to USB

• Port

: Specify the port No. of the USB.

• Baud rate

: 115200 bits/s

• Data length

: 8 bits

• Parity

: Even

• Stop bit length : 1 bit

• When connected to PORT2

• Port

: Specify the port No. of the RS-232C.

• Baud rate

: 9600 bits/s

• Data length

: 7 bits

• Parity

: Even

• Stop bit length : 1 bit

„ AutoID Navigator

Configure “Option” as follows:
• When connected to USB

• Port

: Specify the port No. of the USB.

* If the port settings are correct, other settings are not required.

• When connected to PORT2

• Port

: Specify the port No. of the RS-232C.

• Connection method

: Check DV-90 PC Direct.

• Baud rate

: 115200 bits/s

• Data length

: 8 bits

• Parity

: Even

• Stop bit length

: 1 bit

6

Send the settings from the setup software.

The parallel input/output LEDs on the DV-90 illuminate
one by one in sequence during communication. The
commands being communicated are displayed on the
LCD.

7

Press the

key on the DV-90 to exit “PC

Direct”.

The settings are completed.

Setting Procedure Using "Quick Setup Code"(DV-90 SetUp Software)

The settings of the code reader can also be configured by reading the barcode on the “Code
Reader "Quick Setup Code" printed from the DV-90 SetUp Software (BL-700/600/500/180).
Refer to 7-31.

Note

If the settings cannot be sent due to the processing speed of the computer, change the
timeout to its maximum value (60 sec).

12

11

10

9

8

7

6

5

4

TRG1

TRG2

3

2

1

<

S E N D

>

O K

ESC

2-2 Setting Procedure of the BL Series and the SR Series