Installation and commissioning examples
Example using DHCP IP address assignment
114
Operating Instructions
10.1.10
Power-up the modules
Power-up
Press the On (supply) side of the power switch to power-up the system. Refer to
"Initial power-up", page 33.
10.1.11
Configure the network with DHCP
The power LED's on each module light solid green. Network LED's initially flash green
five times per second when waiting for DHCP to assign an IP address.
Wait for the Network LED to flash green twice per second on all modules.
Open the
Configuration screen
Click the «Config» button in MSEsetup.
Enable Supervisor
mode
Click the «Supervisor mode» button.
Enter "95148" into the Password field.
Open the Connect
screen
Click the «Connect» button.
Perform a Broadcast
Select the 255.255.255.0 netmask from the Broadcast netmask drop-down list.
Click the «Broadcast» button.
Click the «Cancel» button to proceed without overwriting the settings.
A message verifying that broadcasting is finished is displayed in the Connect
screen message area.
Switch to Static
addressing
Click the «OK» button in the dialog box when asked to switch to static addressing.
10.1.12
Setup the 1 V
PP
channel
Open the Setup
screen
From the Connect screen, click the «Back» button twice.
Click the «Setup» button.
Select the 1 V
PP
Channel
Click on «X1 (CH1)» under MSE 1184 in the Tree.
Click the «Channel» tab.
Assign the length
gauge to the 1 V
PP
channel
Click and drag the Length Gauge icon to the green square labeled X1. The green
square appears when the length gauge icon is clicked and dragged.
Click the «OK» button.
Refer to "Channel setup", page 83 for additional channel settings options.